This is a demo store. No orders will be fulfilled.

A flexible and wearable three-electrode electrochemical sensing system consisting of a two-in-one enzyme-mimic working electrode

ANALYTICA CHIMICA ACTA [2023]
Yanan Wang, Yali Liu, Xiao Wang, Xiyue Cao, Jianfei Xia, Zonghua Wang
ABSTRACT

In this work, a wearable and flexible three-electrode electrochemical sensing system (TESS) by using a two-in-one enzyme-mimic working electrode (TIOWE) is reported. The integrated three-electrode, including working electrodes, reference electrodes, and counter electrodes are formed by transfer printing of Ni 2 P-based composite electrode ink (Ni 2 P/G ink), Ag/AgCl ink, and carbon ink onto PDMS substrate, respectively. The Ni 2 P/G ink-based working electrodes have both good conductivity and enzyme-mimic catalytic activity towards glucose. Under optimized conditions, the TIOWE-TESS has a low detection limit of 0.37 μM and wide linear ranges of 0.001 mM–0.1 mM and 0.1 mM–1.4 mM. Furthermore, the TIOWE-TESS has good applicability in serum samples and reveals remarkable electrochemical performance at fluctuant working temperatures. The proposed TIOWE-TESS can be integrated on a waterproof bandage to fabricate a skin-friendly patch device for sweet glucose monitoring, which highlights its potential applications in flexible and wearable commercial devices for health-monitoring.